IE-NDT DENSY 210 Densitometer

 

A microprocessor based densitometer, which combines very high degrees of accuracy and repeatability with an automated routine for monitoring film processor performance according to EN 584 Part 2. The unit is fitted with an RS 232 C serial interface so that process control data can be downloaded to a computer.

 

Technical data

 

Measuring range:

   0D to 6,00D

Calibration:

   According to ASTM E 1079-00, traceable to

   NIST

Accuracy:

   +/- 0,01D (up to 2,00D)

   +/- 0,02D (above 2,00D)

Repeatability:

   +/- 0,01D

Zero drift:

   +/- 0,02D over 8 hours

Dimensions (L x W x H):

   300 x 320 120mm

Supply required:

   230V AC 50Hz

Weight:

   2,98 kg

 

CAT. NO. 15.15.04

 

 

 

 

 

DIGIT-X Densitometer

 

A hand-held transmission densitometer combining ease of use with high accuracy and repeatability. The fine fibre-optic probe allows small areas of film to be measured accurately with readings being displayed on the large clear LCD panel to 2 decimal places. Battery usage is minimal and low battery warning indicator is provided on the digital display. In use the probe is placed on the illuminated film viewer and the ‘Zero’ button momentarily pressed. The digital display will read 0.00. The instrument is then placed on the film and the density readings obtained.

 

Technical data

 

Measuring range:

   0-4D
Accuracy:

   +/-0.05D

Repeatability:

   +/- 0.02D
Power supply:

   9V PP3 battery
Battery life:

   2,500 hours
Dimensions (Instrument):

   210 x 60 x 40mm
Dimensions (instrument and case):

   235 x 210 x 70mm

Overall weight:

   675 g

 

CAT. NO. 15.15.06

 

 

  

IE- NDT DENSORAPID A Densitometer

 

A compact, battery powered instrument with hand-held "wander" probe for measuring densities while the film is on the viewer. Primarily intended for site use, the unit is supplied complete with mains adaptor/charger in a tough, plastic carrying case.

 

Technical data

 

Measuring range:

   0,1 to 5,0D

Calibration:

   According to ASTM E 1079-00, traceable to

   NIST

Accuracy:

   Better than +/- 0,05D

Film viewer luminance range:

   15.000 to 80.000 Cd/m²

Mains adaptor input voltage:

   230V AC, 50Hz

Carrying case dimensions:

   355 x 275 x 55mm

Overall weight

   1,1 kg

 

CAT. NO. 15.15.05

 

 

 

 

 

 

 

 

 

Transmission density stepwedge film, calibrated to ASTM E 1079

 

Stepwedge film with 14 steps, traceable to NIST, for use as a calibration piece when adjusting and monitoring the performance of densitometers. Supplied in a plastic wallet with certificate.

 

Certificate available in English, French, German, Italian and Spanish.

 

CAT. NO. 15.15.08

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 